Subject: FW: Force Transducer Cal
Date: Tue, 25 Jan 2000 17:02:11 -0600
From: Jim Smith <jim.smith@mts.com>
To: Rod Larson <Rod.Larson@mts.com>

Here's Julie's answer

----------
From: "Grangroth, Julie" <Julie.Grangroth@mts.com>
Date: Tue, 25 Jan 2000 12:14:58 -0600
To: "Smith, Jim" <Jim.Smith@mts.com>
Subject: RE: Force Transducer Cal

Jim/Rod,
I was under the understanding that the original sensor
cartridge may have been lost with the calibration info
on it-if that was the case they could take the info from
the calibration sheets and create a sensor database record.
I in no way intended for this to be applicable to a different
type of electronics being used as a readout device.
The reason we need this important information on our calibration
plan such as the type of electronics being used on the system, is
the parameters vary significantly per types of electronics(readout
devices).
If you have any further questions call me.
Thanks!
Julie
> -----Original Message-----
> From: Jim Smith [SMTP:jim.smith@mts.com]
> Sent: Tuesday, January 25, 2000 11:53 AM
> To: Grangroth, Julie
> Subject: Re: Force Transducer Cal
>
> Julie
>
> This is Rod Larson's response:
>
> Julie in cal lab said that they could load data from diskette, and
> that the
> load cells do not have to come back.  That was at shipment time; has
> she
> changed her mind???
>
> I'm just the messenger here, so I just passed the message along.
>
> Jim Smith
>
> > From: "Grangroth, Julie" <Julie.Grangroth@mts.com>
> > Date: Tue, 25 Jan 2000 08:24:56 -0600
> > To: "Smith, Jim" <Jim.Smith@mts.com>
> > Subject: FW: Force Transducer Cal
> >
> >
> >
> >> -----Original Message-----
> >> From:    Grangroth, Julie
> >> Sent:    Tuesday, January 25, 2000 6:54 AM
> >> To:    Otsuka, Yutaka
> >> Subject:    RE: Force Transducer Cal
> >>
> >> Yutaka,
> >> My calibration plan from Rod Larson/Jim Smith requests
> >> a Teststar II calibration so therefore I calibrated both
> >> transducers to a 490.21 which includes the calibration file
> >> on a disk enclosed in a sensor cartridge.  My suggestion
> >> would be to ship the transducers back here to be calibrated
> >> to Teststar IIs (493.21). I could calibrate the load cells within a
> >> day
> >> of their arrival.  If it is not possible to ship the load cells
> >> back here to Minneapolis let me know.  Teststar IIs calibrations
> >> require a transducer I.D. cable which includes the calibration
> >> file inside the actual cable.  I am sorry about the confusion in
> >> regards to the calibration.  I hope to hear from you soon.
> >> Thanks!
> >> Julie
> >>
> >> -----Original Message-----
> >> From:    Yutaka.otsuka [SMTP:Yutaka.otsuka@mts.com]
> >> Sent:    Tuesday, January 25, 2000 5:00 AM
> >> To:    Grangroth, Julie
> >> Subject:    Re: Force Transducer Cal
> >> Importance:    High
> >>
> >> Hello Julie,
> >>
> >> Thank you for your reply. But we noticed other things.
> >> We have Test StarIIs program on site. but calibration disk come
> with
> >> sensor
> >> cartridge.
> >> Isn't it for Test Star instead of IIs? If that so, do you have any
> >> idea to
> >> convert the cal file?
> >> MTS-Japan has only 500kN master load cell, we can't calibrate the
> >> 1000kN
> >> load cell on site.
> >> Please give some advice.
> >>
> >>
> >> Thanks
> >> Yutaka
> >>
> >>
> >> "Grangroth, Julie" wrote:
> >>
> >>> Yutaka,
> >>> We are not able to send an electronic copy to you
> >>> via a disk.  The problem with an electronic copy is that
> >>> we have lost the source code for the executable file
> >>> which allows us to generate this report.  Gary Dahlberg
> >>> is currently working on generating a new type of report.
> >>> We have sent  electronic copies in the past but
> >>> field service engineers were unable to retrieve any of the
> >>> information that was copied on the disk.  Do you have
> >>> a fax number where I can send the new report with the
> >>> graph on it?  Please let me know where to send this.
> >>> I hope all is well with you.  I hope to hear from you soon.
> >>> Thanks!!!
> >>> Julie
> >>>
> >>>> -----Original Message-----
> >>>> From: Yutaka.otsuka [SMTP:Yutaka.otsuka@mts.com]
> >>>> Sent: Monday, January 24, 2000 5:38 AM
> >>>> To:   Grangroth, Julie; Imafuku, Nobuhiro
> >>>> Subject:      Force Transducer Cal
> >>>> Importance:   High
> >>>>
> >>>> Hello Julie
> >>>>
> >>>>
> >>>> How are you doing? I'm Yutaka. Do you remember me? If you are
> not,
> >>>> please ask Don Johnson.
> >>>> Now I'm in Takenaka site.
> >>>> I got Force Transducer Calibration data MTS ID 11641.
> >>>> One sheet with Graph, and other one is not. I need both with
> Graph
> >> in
> >>>> the sheet.
> >>>> If you made by Micro Soft Excel, could you send me a sheet? I can
> >> fix
> >>>> it
> >>>> by my self.
> >>>>
> >>>> Thank you for your kindness.
> >>>>
> >>>> Yutaka. Otsuka
> >>>> mailto:Yutaka.Otsuka@mts.com
> >>>>