Subject: RE: Force Transducer Cal
Date: Wed, 26 Jan 2000 15:15:12 -0600
From: "Cross, Dan" <Dan.Cross@mts.com>
To: "Smith, Jim" <Jim.Smith@mts.com>, "Larson, Rod" <Rod.Larson@mts.com>

Jim, Rod

I talked briefly with Barry Rice about this.
The only way we could come up with to do this is to match the shunt cal
values from the TestStar II calibration.  The problem isn't with getting
values into the Transducer ID EERom, the TestStar IIs software can do
this.  The problem is that you don't have accurate enough values for
excitation and gain that you can use to transfer to the TestStar IIs.

Teststar II has a different shunt cal resistor for each range whereas
Teststar IIs has a single shunt resistor.  To transfer the cal for a
range you will need to temporarily move the proper shunt resistor from
the Teststar II cartridge into the Teststar IIs Transducer ID box, set
the excitation to 10.0v and adjust the gain to match the shunt offset
value recorded for that range on the TestStar II cal sheet.

If you had a TestStar II you would have had to adjust the excitation
voltage while performing a shunt cal in Range1 inorder to correct for
differences between the conditioner used by the cal lab and the customer
conditioner anyway.  This procedure is not too much different.

When you have adjusted all the ranges I would then put the original
shunt resistor back into the Tranducer ID box and record the new shunt
offset values for each range.

Maybe the cal lab could run a test to verify the accuracy of transfering
the calibration in this way.

Dan

-----Original Message-----
From: Jim Smith [mailto:jim.smith@mts.com]
Sent: Wednesday, January 26, 2000 8:15 AM
To: Cross, Dan
Subject: FW: Force Transducer Cal

Dan
Rod Larson asked me to touch base with you.  We had 2 load cells
calibrated
for a job in Japan for "Teststar II" and they should have been Teststar
IIs.
Per Rod's comments below, is there any way to do this in the field.  It
will
cost us in the neighborhood of $10,000 to ship them back from Japan!

Julie in the Cal Lab has the calibration spec sheets if you need them.

Please let me know or just reply to Rod.Larson@mts.com.  He is in Japan,
but
he does check his mail often.

Thanks

Jim Smith

----------
From: Rod Larson <rod.larson@mts.com>
Date: Tue, 25 Jan 2000 19:53:03 -0600
To: Jim Smith <jim.smith@mts.com>
Subject: Re: FW: Force Transducer Cal

Jim,
I have the correct kind of cable for the Cobra system; just the
calibration
data is NOT.  I have the right kind of "funny" cable to hook up to the
load
cell; the calibration data is just NOT loaded on the f*king thing.
Julie
told me that the data could be loaded in the right cable.  Yes, the
calibration was done wrong cuz I screwed up; however, I got the right
kind
of Cobra cable from Chaska and then was told that the calibration data
can
be loaded into the Cobra cable pod.  I cannot believe that we have to
spend
$10,000 to send those load cells back to MTS to load 2 numbers in the
damn
conditioners.

Ask Dan Cross (TIIs S/W guy) about this problem; I cannot believe this.
Rod

At 05:02 PM 1/25/00 -0600, you wrote:
>Here's Julie's answer
>
>----------
>From: "Grangroth, Julie" <Julie.Grangroth@mts.com>
>Date: Tue, 25 Jan 2000 12:14:58 -0600
>To: "Smith, Jim" <Jim.Smith@mts.com>
>Subject: RE: Force Transducer Cal
>
>Jim/Rod,
>I was under the understanding that the original sensor
>cartridge may have been lost with the calibration info
>on it-if that was the case they could take the info from
>the calibration sheets and create a sensor database record.
>I in no way intended for this to be applicable to a different
>type of electronics being used as a readout device.
>The reason we need this important information on our calibration
>plan such as the type of electronics being used on the system, is
>the parameters vary significantly per types of electronics(readout
>devices).
>If you have any further questions call me.
>Thanks!
>Julie
>> -----Original Message-----
>> From: Jim Smith [SMTP:jim.smith@mts.com]
>> Sent: Tuesday, January 25, 2000 11:53 AM
>> To: Grangroth, Julie
>> Subject: Re: Force Transducer Cal
>>
>> Julie
>>
>> This is Rod Larson's response:
>>
>> Julie in cal lab said that they could load data from diskette, and
>> that the
>> load cells do not have to come back.  That was at shipment time; has
>> she
>> changed her mind???
>>
>> I'm just the messenger here, so I just passed the message along.
>>
>> Jim Smith
>>
>> > From: "Grangroth, Julie" <Julie.Grangroth@mts.com>
>> > Date: Tue, 25 Jan 2000 08:24:56 -0600
>> > To: "Smith, Jim" <Jim.Smith@mts.com>
>> > Subject: FW: Force Transducer Cal
>> >
>> >
>> >
>> >> -----Original Message-----
>> >> From:    Grangroth, Julie
>> >> Sent:    Tuesday, January 25, 2000 6:54 AM
>> >> To:    Otsuka, Yutaka
>> >> Subject:    RE: Force Transducer Cal
>> >>
>> >> Yutaka,
>> >> My calibration plan from Rod Larson/Jim Smith requests
>> >> a Teststar II calibration so therefore I calibrated both
>> >> transducers to a 490.21 which includes the calibration file
>> >> on a disk enclosed in a sensor cartridge.  My suggestion
>> >> would be to ship the transducers back here to be calibrated
>> >> to Teststar IIs (493.21). I could calibrate the load cells within
a
>> >> day
>> >> of their arrival.  If it is not possible to ship the load cells
>> >> back here to Minneapolis let me know.  Teststar IIs calibrations
>> >> require a transducer I.D. cable which includes the calibration
>> >> file inside the actual cable.  I am sorry about the confusion in
>> >> regards to the calibration.  I hope to hear from you soon.
>> >> Thanks!
>> >> Julie
>> >>
>> >> -----Original Message-----
>> >> From:    Yutaka.otsuka [SMTP:Yutaka.otsuka@mts.com]
>> >> Sent:    Tuesday, January 25, 2000 5:00 AM
>> >> To:    Grangroth, Julie
>> >> Subject:    Re: Force Transducer Cal
>> >> Importance:    High
>> >>
>> >> Hello Julie,
>> >>
>> >> Thank you for your reply. But we noticed other things.
>> >> We have Test StarIIs program on site. but calibration disk come
>> with
>> >> sensor
>> >> cartridge.
>> >> Isn't it for Test Star instead of IIs? If that so, do you have any
>> >> idea to
>> >> convert the cal file?
>> >> MTS-Japan has only 500kN master load cell, we can't calibrate the
>> >> 1000kN
>> >> load cell on site.
>> >> Please give some advice.
>> >>
>> >>
>> >> Thanks
>> >> Yutaka
>> >>
>> >>
>> >> "Grangroth, Julie" wrote:
>> >>
>> >>> Yutaka,
>> >>> We are not able to send an electronic copy to you
>> >>> via a disk.  The problem with an electronic copy is that
>> >>> we have lost the source code for the executable file
>> >>> which allows us to generate this report.  Gary Dahlberg
>> >>> is currently working on generating a new type of report.
>> >>> We have sent  electronic copies in the past but
>> >>> field service engineers were unable to retrieve any of the
>> >>> information that was copied on the disk.  Do you have
>> >>> a fax number where I can send the new report with the
>> >>> graph on it?  Please let me know where to send this.
>> >>> I hope all is well with you.  I hope to hear from you soon.
>> >>> Thanks!!!
>> >>> Julie
>> >>>
>> >>>> -----Original Message-----
>> >>>> From: Yutaka.otsuka [SMTP:Yutaka.otsuka@mts.com]
>> >>>> Sent: Monday, January 24, 2000 5:38 AM
>> >>>> To:   Grangroth, Julie; Imafuku, Nobuhiro
>> >>>> Subject:      Force Transducer Cal
>> >>>> Importance:   High
>> >>>>
>> >>>> Hello Julie
>> >>>>
>> >>>>
>> >>>> How are you doing? I'm Yutaka. Do you remember me? If you are
>> not,
>> >>>> please ask Don Johnson.
>> >>>> Now I'm in Takenaka site.
>> >>>> I got Force Transducer Calibration data MTS ID 11641.
>> >>>> One sheet with Graph, and other one is not. I need both with
>> Graph
>> >> in
>> >>>> the sheet.
>> >>>> If you made by Micro Soft Excel, could you send me a sheet? I
can
>> >> fix
>> >>>> it
>> >>>> by my self.
>> >>>>
>> >>>> Thank you for your kindness.
>> >>>>
>> >>>> Yutaka. Otsuka
>> >>>> mailto:Yutaka.Otsuka@mts.com
>> >>>>
>