Yes, TSIIs chassis light go off. I was looking to use connect the debug cable
to boot the Vxworks(Chassis program). It boot up the correctly.

And as I heard, some Flex IIm or TSIIs suddenly lost the communication, when
use the on board ether net compaq PC. Then they changed the communication
through the 3com and it fixed. So, I want to try today.
I will let you know. How it works.

Yutaka

Rod Larson wrote:

> IT is NOT good that 2 different voltages; however, I am not sure which is
> correct.  But, cannot have 2 different.  Are both on SAME power circuit.
> The TestStar IIs box is auto switching from 100VAC to 240VAC input.  The
> digital i/o box should have a 24VDC power supply in it.  This power supply
> is ONLY used  for OPTO-22 power and that would not cause a problem.
>
> You MUST interface to the TSIIs box w/ COMPAQ ethernet card whose
> properties are set to:
> HALF DUPLEX
> 100 MBps
>
> If this is not the case, then it may not boot.  Does the light go off on
> the TSIIs chassis when you power up the TSIIs box (meaning that the driver
> s/w loaded on power up correctly.
>
> Rod Larson
>
> At 01:09 AM 3/1/00 +0900, you wrote:
> >Hello Guys,
> >
> >How are you?
> >
> >I still can't boot the system. But I found one thing,
> >
> >Rod-san,
> >
> >Analog chassis power supply voltage are different. 100 ton voltage 15
> >Volts, 50 ton voltage has 24 Volts.
> >Why those are different. Which voltage is correct or supposed to be? In
> >Tokyo office, I got only 15 Volts power supply.
> >
> >I heard some of other TSIIs and FlexIIm had communication problem,
> >especially use compaq PC.
> >They solved that to use 3com ether net card, I will try tomorrow.
> >
> >I hope I can give you good news tomorrow.
> >
> >Yutaka
> >
> >