Subject: Re: TestStar IIs S/W Patch
Date: Mon, 28 Feb 2000 00:18:12 +0900
From: "Yutaka.Otsuka" <Yutaka.Otsuka@mts.com>
To: Rod Larson <rod.larson@mts.com>, rodl@scc.net

Hello Rod-san,

How are you?
Do I need to put this software patch to fix 469DU?
If it's OK to put later, I will do after fix the 469DU.
Please let me know.

RGDS,
Yutaka

Rod Larson wrote:

> Fuku, Yutaka,
> The TestStar IIs Software has a bug:
>
> If you do a control mode-switch with the hydraulics OFF; then the ramp
> generator is using the WRONG units.  If you mode-switch w/ the
> hydraulics ON, then everything is OK.  The included file patches the s/w
> to correct the bug.  A new version will be sent to you as soon as it has
> been fully tested.
> Rod Larson
>
>   ------------------------------------------------------------------------
>                                                Name: R2.4F Auto-Offset - Modeswitch Patch.exe
>    R2.4F Auto-Offset - Modeswitch Patch.exe    Type: $B%?%$%WL$;XDj(J (application/octet-stream)
>                                            Encoding: base64