18 Hours March 2012 1. 5.0 Hours Change of DTS software to switch the displacement conditioners from AC to DC so we can use analog temposonics for displacement feedback on the DTS2500KN. 2. 2.0 Hours Check on triple power connections on the BTS Temposonics to see if ok to use for DTS2500KN Tempo power. Checked OK and then made sure we had enough current for all Tempos. That's OK too. 3. 1.5 Hour Cable check for rest of the cables for DTS2500KN. All OK. 4. 1.0 Hour Rexroth Valve setup for BTS 493.15 3-stage valve driver. AC excitation = 0.0 Volts. 5. 0.5 Hour How to get out of Simulation mode. April 2012 6. 2.5 Hour assist polarity of signals and calibration of stroke and load of DTS2500KN. 7. 2.0 Hours create calx.doc word document to get stroke calibration right. 8. 0.5 Hour load cell calibration and initial tuning 9. 3.0 Hours mess-up on valve balancing - NOT charged for several days lost on this account. Sum will NOT show these hours. 10. 0.5 Hour correct valve balance adjustment - turn off 1 valve hydraulics at a time. 11. 1.0 Hour explanation of Temposonics vibration sensitivity. May 12. 1.5 Hours running problems with the DTS2500KN. Looking at run data. Total = 21 Hours - 3 Hours = 18 Hours total for DTS2500KN project.